Sistem Pengontrolan Suhu dan Kelembaban pada Inkubator Bayi

Laura A. S. Lapono
Journal article Jurnal Fisika • 2016

Abstract

  Baby incubator is very important to keep the newborn's body temperature especially for prematuture babies. The incubator is designed to have a lenght of 60 cm, a width of 40 cm, and a height of 30 cm. System of baby incubatir will automatically turn on or turn off the fan and or heating in accordance with the normal range of temperature and humidity in the incubator. The normal limits of temperature used is 33°C to 35°C. While the normal limits of air humidity in the incubator used is 40% sampai 60%. Data acquisition system consists of temperature and humidity sensor, microcontroller ATmega8535, fan, heater, and LCD. LCD is used to display the results of measurements of temperature and humidity.Heater is used to regulate the temperature inseide the incubator. While fan is used to regulate the humidity in the incubator. Test results show that the heater will turn on if the temperature is below the limits of 33°C. While the fan will turn on if the humidity is above 60%.   Keywords: baby incubator, temperature and humidity sensor, microcontroller ATmega8535